Dual-Channel Conductivity Analyzer
Model Code | FLXA21 | FLXA202 | DC402G |
---|---|---|---|
Area Classification |
|
|
|
Enclosure Material | Plastic (Polycarbonate) | Aluminum alloy die cast with epoxy coating (Standard offering) Aluminum alloy die cast with urethane or high anti corrosion coating (available upon request) | Cast Aluminum with chemically resistance coating; polycarbonate cover with polycarbonate flexible window |
Housing Rating | IP66/ NEMA 4A/ CSA 3S | IP66 (except Canada), Type 4X (except Canada), Type 3S/4X (Canada) | IP66/ NEMA 4A/ CSA 3S |
Power Supply | 2-wire 24VDC, Loop Powered | 2-wire 24VDC, Loop Powered | 4-Wire 110-256 VAC, Mains Power |
Output Signal | Bi-directional HART digital communication, superimposed on mA (4-20mA) signal. | Bi-directional HART digital communication, superimposed on mA (4-20mA) signal. | Two isolated outputs of 4 to 20 mA DC with common negative. |
The model FLEXA® two-wire analyzer is used for continuous on-line measurements in industrial installations. It offers an option for single or dual sensor measurement, making it the most flexible 2-wire analyzer available. The model FLEXA® modular-designed series analyzer.
One analyzer can accept any of 4 types of measurements; pH/ORP, Contacting Conductivity (SC), Inductive Conductivity (ISC) and Dissolved Oxygen (DO)
Dual sensor measurement on a 2-wire type analyzer pH/ORP and pH/ORP, SC and SC, and DO and DO
Modular design: replaceable sensor modules
Redundant system on dual sensor measurement
Variety of materials of construction: SS or Plastic
Intuitive easy touch screen operation on 2-wire type analyzer
Unique HMI menu structure in 12 languages
Quick setup menu for fast and east measurement operation
Online Sensor Wellness checking for predictive maintenance
NEMA 4X / IP66 Enclosure
Flexibility, reliability and low maintenance are among the benefits provided by the EXA DC402G conductivity analyzer. Designed to meet the exacting requirements of measuring dual cell conductivity and resistivity in the modern industrial environment, it contains many features to ensure the best precision whatever the application. Differential, ratio, deviation. % passage or % rejection can be indicated and/or transmitted.
This 4-wire converter is housed in a robust IP65 field mountable case. Two mA outputs, four relays, digital communication and a clear LCD make the DC402G a truly comprehensive package.
The DC402G features PI control on the auxilliary mA output and the pulse proportional relay outputs, thus avoiding the need for a separate controller.
The famous EXA sensor diagnostics are also present in the DC402G. Self-tuning of the excitation frequency and measuring pulses ensure optimum accuracy. The polarisation check gives on line indication of sensor fouling and early warning that maintenance is needed. A wide variety of temperature compensation possibilities (NaCl according to IEC 746-3, manual TC, preprogrammed matrices and a freely programmable 5x5 matrix), provides a high-accuracy measurement with minimum effort.
Differential, ratio or calculated outputs
Universal conductivity/resistivity, software switchable
On-line sensor checking
Self-tuning measuring signal
Matrix temperature compensation for pure water applications
Four fully configurable SPDT contact outputs
Two fully configurable mA outputs
Built-in PI controller
Easy to use EXA control panel
USP 23 monitoring